We have outlined our progress with respect to developing a novel device for monitoring oral samples for bacterial and/or viral pathogens. The system is based on an existing device for measuring drugs of abuse in an oral sample. The sample is collected on an absorbent pad that delivers a metered dose to the cassette. The sample is then separated into 4 channels for the detection of antigen, RNA or DNA, and host antibodies to the pathogen. The detection system involves the Upconverting Phosphor Technology (UPT), whereby the captured pathogen analyte is detected by interrogation of the UPT particles with near-infrared light, and the emitted visible light is detected by the analyzer. Several of the steps in this process have already been worked out for viral and/or bacterial pathogens, and most of the remaining effort will be aimed at integrating these steps into a single microfluidic device while maintaining the current sensitivity.

Advisor

Date Range for Data Collection (Start Date)

Date Range for Data Collection (End Date)

Digital Object Identifier

Series name and number

Publication date

2005-06-01

Journal title

Volume number

Issue number

Publisher

Publisher DOI

Journal Issues

Comments

Reprinted by permission. Reprinted in Advances in Dental Research, Volume 18, June 2005, pages 12-16. Publisher URL: http://adr.iadrjournals.org/

Recommended citation

Collection